Tutorial: Introduction to Aperture
Photometry
This tutorial describes the procedure for making a
simple photometric measurement using Mira's
Aperture Photometry package. This tutorial assumes
you understand how to display an image and do elementary operations
like moving the image cursor. To get a quick start on displaying
images, see
Tutorial: Displaying an Image. A separate
tutorial describes how to do
Time Series Photometry. One application area
for time series photometry is finding the time of minimum
brightness for variable stars and other objects; for this
application, see the
Kwee - van Woerden Solver command.
Getting Started
To begin, click File > Open on the menu
to open the familiar Open dialog. In the Open dialog,
select the image shown below and click [Open].
The image opens as shown below, with the
Image Bar at the top of the window. The yellow
rectangle shows the part of the image used in this tutorial.
To get more space without showing the
Image Bar at the top of the window, right click on
the image and select Configure > Image
Bar from the menu to close the Image Bar (you can re-open it
the same way). Adjust the position and magnification to show the
field of view below (for help with magnifying, see
Magnifying the Image). On the main
Image Measurements Toolbar, click the
photometry button to open the
Aperture Photometry Toolbar on the left border
of the image window.
Before measuring a star, we want to magnify a
portion of the field so that it looks like the view shown below
Now click on the main
toolbar. This displays the Aperture Photometry Toolbar on
the left border of the window (see picture below).
Preparing the Image for Photometry
In this tutorial, we are measuring a single image,
but Mira can measure any number of images displayed as an
image set. Before starting photometry, it is
important to know that each image contains information needed for
calculating magnitudes and their errors (uncertainties). These
quantities, stored in the image header, are the exposure time,
electronic gain, and readout noise, and they are stored as the
keywordsEXPTIME, GAIN, and RDNOISE.
If these keywords are not in each image header, or the values are
not correct, then you need to add or edit them. Mira provides a way
to do this efficiently, even for many images in an image set; see
Fixing Header Problems in Photometric Data. You
also can have Mira substitute aliases for the standard keywords;
see
Photometry Keywords.
Making the Measurement
Now move the mouse pointer to near coordinate
(160,458) and click the left mouse button. Mira computes a centroid
position near this point and draws a marker on the star near where
you clicked:
The magnitude is reported with other information in
the Photometric Measurements
Report Window shown below. This Report Window lists
more than 20 columns of data but we see only the first few in this
example. To show more, you can adjust the data table using the
scrollbar. See
Photometric Measurement Definitions to learn
more about this these quantities.
Adding a Standard Star
In the Photometric Measurements window shown above,
notice that the magnitude is a large negative number. This results
from the definition of magnitude as -2.5 log (flux) and the fact
that a photometric zero point has not yet been applied (see
Aperture Photometry). We will use a standard star to
calculate the zero point. On the Aperture Photometry Toolbar, click
to enter standard star marking mode,
then click on the standard star. In this case, move the mouse onto
the star near coordinate (57,453) and click to mark the object.
This opens the Standard Data dialog:
Into the Standard Data dialog, enter the
star's magnitude and a statistical weight. You can leave the weight
at the default of 1. If you are doing "ensemble photometry", you
may wish to use another weight value based on the magnitude or the
magnitude uncertainty of the standard star. For the star you just
marked, enter 12.5 for the magnitude and click [OK] to accept the value. The image window
updates like this:
Notice that the standard star has an ID number with
an asterisk (*) beside it; this is used in the photometry report to
show that the measurement was made of a standard star. If the
magnitudes in the table are still negative after marking a standard
star, click to compute and apply the
photometric zero point. The table now has positive magnitudes.
Notice that the standard star you marked has the magnitude 12.5000
that you entered:
To see more data for these measurements, use the
horizontal scrollbar. Two
photometric error estimates are listed in the
report: Error and Error(T). These are the empirical (measured)
uncertainty and the theoretical uncertainty for the calculated
magnitude. Mira also saves the photometric zero point and its
uncertainty to the image header (see ZERO-PT and ZERPTERR
keywords in the table of
FITS Keywords), but these values are not saved
to files unless you save the images using
File > Save or an equivalent command. With
the zero point saved in the image header, you can do photometry on
another occasion and the magnitude measurements will automatically
be convert raw to instrumental values with the zero point
applied.
Measuring More Objects
Now let us return to measuring magnitudes. Click
on the toolbar to switch to
Marking Mode. Click 3 more stars as
shown below, and then switch to Standard mode. Mark object 6 as a
second standard star and enter the value 12.62 for its magnitude,
then click [OK] to close the
dialog.
Now click to recalculate the
zero point using 2 standards. This produces the table below. Notice
that the two standard stars now have magnitudes that are not the
same as the values you entered. This happens because their
magnitudes are calculated using the zero point value. Each standard
star now has a residual, which is the difference between its
calculated magnitude and its predicted magnitude based on the
photometric zero point. The residuals are listed in one of the
columns to the right of the table (you will have to use the
horizontal scrollbar to view it).
Now that there are 6 measurements in the report
table, you may wish to save the se results to a file or export them
to other software such as Microsoft Excel where you might do
further analysis. This simple procedure is described in
Grid Controls. If you save the coordinates to a file,
you can use the Import command on the
Point Markers page to load the object
coordinates at a future date.
By now you have doubtless noticed the scrolling
Text Editor Window named Photometry Messages.
This window lists results of photometry calculations and gives
warning messages. The contents of the message window can be edited,
printed, copied, or saved to a file.
Changing the Apertures
Suppose we did not like the aperture sizes and
shapes that were used for these measurements. These can be changed
using the
Aperture Tool Window. To open the Aperture Tool,
click on the Aperture Photometry Toolbar.
This tool window is a shared resource used by whichever photometry
window is currently active). When you change the aperture
description, Mira automatically re-measures and updates the report
table.
Going Further
Mira can perform other useful operations that will
enhance your scientific capabilities. Here are some examples:
-
Find the time of minimum brightness for a
variable star or similar object using the
Kwee - van Woerden Solver.
-
Use
File Event Scripting to do real-time monitoring,
processing, and/or magnitude extraction from images are they are
saved.
-
Perform "time series photometry" by tracking,
measuring, and calibrating objects through an
image set.
-
Copy and paste object markers from an image
window with active Aperture Photometry window to another
image window.
-
Import data from a
catalog database by (column,row) or (RA, Dec).
The objects can be tracked to other images, measured, and
calibrated for the entire image set.
Related Topics
Contents
Tutorials
Getting Started
Aperture Photometry
Tutorial: Doing Time Series Photometry
Plot Light Curve
AAVSO Report
Fixing Header Problems in Photometric Data
FITS Format
Mira Pro x64 User's Guide, Copyright Ⓒ 2023 Mirametrics, Inc. All
Rights Reserved.
|